![Kendro HERAcell 150 Operating Instructions Manual Download Page 130](http://html1.mh-extra.com/html/kendro/heracell-150/heracell-150_operating-instructions-manual_663805130.webp)
Name/Name
Datum/Date
Benennung/Designation
Erstellt/Prepared
D.Dornseiff
01/15/98
Index
Änderung
Name/Name
Datum/Date
Service Handbuch HERAcell 150
Service Manual HERAcell 150
C
202613
D.Dornseiff
06/26/03
Dokumentnr./Document No.
Seite/Page
Datei/File
50051109_04_EN_C.doc
50051109 / C
24/ 9
Weitergabe sowie Vervielfältigung dieser Unterlage, Verwertung oder Mitteilung ihres Inhalts nicht gestattet, soweit nicht ausdrücklich zugestanden. Zuwiderhandlungen ver-
pflichten zu Schadensersatz. Alle Rechte für den Fall der Patenterteilung oder Gebrauchsmustereintragung vorbehalten. © Kendro Laboratory Products GmbH, Germany
Multimeter
Check the power supply
to the O2 board and
flask rotating equipment.
Replace the mains power
supply unit.
55
Bus cable and
sensor block
both in working
order.
Switch the unit off.
Disconnect the bus ca-
ble to the sensor block
at the main board.
Attach the test set (ca-
ble/sensor block).
Switch the unit on.
Perform the functional
test.
Test set operational:
Repeat the functional test of
the individual installed com-
ponents. Replace the cable
or sensor block.
Test set not operational:
Replace the main board.
99
Ohmmeter
The door switch contact
is closed when the glass
door is open!
Perform the functional
test at the main bard in-
put.
If defective: Replace the
door switch.
If defective: Replace the
main board.
77
Error in the CO
2
cal range:
Initiate an auto-start.
Temperature adjustment er-
ror:
If the error reoccurs, re-
place the measuring cell.
Replace the measuring
cell.
88
Check the setup loca-
tion (drafty, direct sun-
light, etc.)
If necessary, change the
setup location/conditions.
Replace the measuring
cell.
Summary of Contents for HERAcell 150
Page 82: ...CO2 Incubator HERAcell 150 Operating Instructions 82 16 Certificate of decontamination...
Page 83: ...CO2 Incubator HERAcell 150 Operating Instructions 83 16 Certificate of decontamination...
Page 84: ...Internet http www kendro com...
Page 85: ......
Page 87: ...Valid 10 2005 50047292 B Gas Cylinder Change Over Unit GM 2 Operating instructions GB...
Page 106: ......
Page 152: ......
Page 153: ......
Page 154: ......
Page 155: ......
Page 156: ......
Page 157: ......
Page 158: ......
Page 159: ......
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 237: ...Version 04 2004 50078754 Accessories for CO2 incubators Illustrations and describtion...
Page 249: ...Footprint HERAcell...
Page 250: ...Footprint HERAcell...