CO
2
Incubator HERAcell
®
150 Operating Instructions
28
Fig. 12
Rear panel openings
4.
Description of the device
Rear panel openings:
Fig. 12:
A sealable access port [1] allows cables, hoses or ad-
ditional sensors to be routed into the work space of the device.
A pressure compensation opening with insert [2] at the rear
panel of the device allows a compensation between the pres-
sures in the work space and in the operating room.
NOTE – Operating conditions
When accessories are to be operated in the
work space of the CO
2
incubator, the ambient
condition requirements must be observed (see
table below). The energy introduced into the
work space affects the beginning of the tem-
perature control range. When additional hea-
ting sources are introduced into the work
space, condensation (e.g. at the glass door)
may occur.
Introduced energy
0 W
RT + 3° C
24° C
5 W
RT + 6,5° C
27,5° C
10 W
RT + 9,5° C
30,5° C
15 W
RT + 13° C
34° C
20 W
RT + 16° C
37° C
Control range of the temperature
*RT = Ambient temperature
General
Example:
RT* = 21° C
Summary of Contents for HERAcell 150
Page 82: ...CO2 Incubator HERAcell 150 Operating Instructions 82 16 Certificate of decontamination...
Page 83: ...CO2 Incubator HERAcell 150 Operating Instructions 83 16 Certificate of decontamination...
Page 84: ...Internet http www kendro com...
Page 85: ......
Page 87: ...Valid 10 2005 50047292 B Gas Cylinder Change Over Unit GM 2 Operating instructions GB...
Page 106: ......
Page 152: ......
Page 153: ......
Page 154: ......
Page 155: ......
Page 156: ......
Page 157: ......
Page 158: ......
Page 159: ......
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 237: ...Version 04 2004 50078754 Accessories for CO2 incubators Illustrations and describtion...
Page 249: ...Footprint HERAcell...
Page 250: ...Footprint HERAcell...