Name/Name
Datum/Date
Benennung/Designation
Erstellt/Prepared
D.Dornseiff
01/15/98
Index
Änderung
Name/Name
Datum/Date
Service Handbuch HERAcell 150
Service Manual HERAcell 150
C
202613
D.Dornseiff
06/26/03
Dokumentnr./Document No.
Seite/Page
Datei/File
50051109_04_EN_C.doc
50051109 / C
14/ 9
Weitergabe sowie Vervielfältigung dieser Unterlage, Verwertung oder Mitteilung ihres Inhalts nicht gestattet, soweit nicht ausdrücklich zugestanden. Zuwiderhandlungen ver-
pflichten zu Schadensersatz. Alle Rechte für den Fall der Patenterteilung oder Gebrauchsmustereintragung vorbehalten. © Kendro Laboratory Products GmbH, Germany
2
Adjustments for regulating range 2 (decontamination mode, 0 - 100 °C)
CAUTION: Specified set value = 90 °C
FL position
FL value range
Description
1
0 = (+) / -0 = (-)
Temperature offset value sign
2
0 ..... 99
Decontamination temperature offset
value, in °C
3
0. ..... 1
Decontamination temperature amplifica-
tion, pre-decimal places
4
000 ..... 999
Decontamination temperature amplifica-
tion, decimal places
5
55 ..... 90
Decontamination temperature set value
(90°C)
6
30 ..... 999
Decontamination hold time, in min. (total
time = equilibrium time + time at 90°C *)
* This holding time begins when the unit reaches the lower limit band value for the
90°C routine, that is, 85°C. The specified set value of 540 min. may not be changed!
3 Outputs, manual control (main PCB)
FL position
FL value range
Description
1
0 / 1
Heater actuator, unit body: X5...X8
2
0 / 1
Heater actuator, unit door: X3
3
0 / 1
Heater actuator, base: X10
4
0 / 1
Actuator, CO
2
valve: JP5
5
0 / 1
Actuator, common alarm: X11
6
0 / 1
Actuator, horn
7
0 / 1
Actuator, fan
8
0 / 1
Actuators remain active as long as the
cal
key is pressed.
9
0 / 1
O
2
valve switch
The output actuators can be manually activated on this function level. The switch state
(1) can be reset by:
- Manually returning it to 0, or;
- Performing a mains reset.
Summary of Contents for HERAcell 150
Page 82: ...CO2 Incubator HERAcell 150 Operating Instructions 82 16 Certificate of decontamination...
Page 83: ...CO2 Incubator HERAcell 150 Operating Instructions 83 16 Certificate of decontamination...
Page 84: ...Internet http www kendro com...
Page 85: ......
Page 87: ...Valid 10 2005 50047292 B Gas Cylinder Change Over Unit GM 2 Operating instructions GB...
Page 106: ......
Page 152: ......
Page 153: ......
Page 154: ......
Page 155: ......
Page 156: ......
Page 157: ......
Page 158: ......
Page 159: ......
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 237: ...Version 04 2004 50078754 Accessories for CO2 incubators Illustrations and describtion...
Page 249: ...Footprint HERAcell...
Page 250: ...Footprint HERAcell...