CO
2
Incubator HERAcell
®
150 Operating Instructions
56
7.
Operation
7.2
Starting operation
Starting and loading the device:
1. Fill the water tray with up to max. 3.0 l of processed water. Do not exceed
the upper level mark.
2. Make sure that the CO
2
/N
2
/O
2
supply system valves are open.
3. Turn the device on using the power switch.
4. Set nominal values for temperature and CO
2
/ O
2
content at the operating
panel.
5. Ventilate work space by leaving both device doors open until acoustic alarm
sounds.
6. Start device using auto-start routine.
7. Close device doors.
8. The temperature control adjusts the temperature to the set nominal value,
humidity rises.
9. When temperature and relative humidity are constant, the automatic adju-
stment of the CO
2
/O
2
measuring system is performed.
10.The ”auto-start” indicator goes off.
11.The CO
2
/O
2
control supplies the set amount of CO
2
/O
2
.
12.The device is ready for operation.
13.Load work space with cultures.
NOTE – Duration of the auto-start routine
When the device is cold and when the ambi-
ent temperature is low, the auto-start routine
may take up to 10 hours.
NOTE – Charge
To ensure sufficient air circulation and even
heating of the samples, the charge surface
within the work space should be used up to 70
% max. Voluminous objects in the work spa-
ce that dissipate heat may impair heat
distribution.
Summary of Contents for HERAcell 150
Page 82: ...CO2 Incubator HERAcell 150 Operating Instructions 82 16 Certificate of decontamination...
Page 83: ...CO2 Incubator HERAcell 150 Operating Instructions 83 16 Certificate of decontamination...
Page 84: ...Internet http www kendro com...
Page 85: ......
Page 87: ...Valid 10 2005 50047292 B Gas Cylinder Change Over Unit GM 2 Operating instructions GB...
Page 106: ......
Page 152: ......
Page 153: ......
Page 154: ......
Page 155: ......
Page 156: ......
Page 157: ......
Page 158: ......
Page 159: ......
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 237: ...Version 04 2004 50078754 Accessories for CO2 incubators Illustrations and describtion...
Page 249: ...Footprint HERAcell...
Page 250: ...Footprint HERAcell...