Use the "cal" key to perform the zero-point adjustment of the CO
2
metering circuit.
Run a test gas addition to check the metering and regulating functions with CO
2
.
For O
2
/N
2
units:
-
Remove the CO
2
from the unit.
-
Run a test gas addition with 70% O
2
or 10% N
2
.
-
Observe the CO
2
zero point.
-
If required, start the O
2
compensation (see Sec. 14.4.3).
Remove the gas from the unit and restart the unit with the -auto-start- routine.
Unit with IR measuring cell:
Se Section 4.14.5.
4.12 MAIN BOARD REPLACEMENT
After completing the service tasks:
After the main board has been replaced and all functional tests and control
measurements have been performed, the unit MUST be restarted with the
auto-start routine.
The unit must be disconnected from the mains power.
Replace the PCB.
The PC bus lines can be connected to the sockets in any desired order.
After installation, switch the unit on.
Press and hold down the -cal- key.
This will reset any -cal- adjustments that may have been performed.
The unit is now operating with its original specified adjustment values.
In units equipped with copper interior fittings, set the door factor in FL 8. Refer to
Section 4.5, Control and Regulating System Configuration.
The replacement PCB is set up for units with stainless steel interior fittings. (Door
factor for copper interior fittings 2.2).
Test the unit’s control and regulating functions:
-
Allow the unit to heat up to, for example, 37.0°C.
-
Once the unit reaches the specified temperature, use the -cal- key to perform
Summary of Contents for HERAcell 150
Page 82: ...CO2 Incubator HERAcell 150 Operating Instructions 82 16 Certificate of decontamination...
Page 83: ...CO2 Incubator HERAcell 150 Operating Instructions 83 16 Certificate of decontamination...
Page 84: ...Internet http www kendro com...
Page 85: ......
Page 87: ...Valid 10 2005 50047292 B Gas Cylinder Change Over Unit GM 2 Operating instructions GB...
Page 106: ......
Page 152: ......
Page 153: ......
Page 154: ......
Page 155: ......
Page 156: ......
Page 157: ......
Page 158: ......
Page 159: ......
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 237: ...Version 04 2004 50078754 Accessories for CO2 incubators Illustrations and describtion...
Page 249: ...Footprint HERAcell...
Page 250: ...Footprint HERAcell...