Service Handbuch HERAcell 150
Service Manual HERAcell 150
C
202613
D.Dornseiff
06/26/03
Dokumentnr./Document No.
Seite/Page
Datei/File
50051109_04_EN_C.doc
50051109 / C
21/ 9
Weitergabe sowie Vervielfältigung dieser Unterlage, Verwertung oder Mitteilung ihres Inhalts nicht gestattet, soweit nicht ausdrücklich zugestanden. Zuwiderhandlungen ver-
pflichten zu Schadensersatz. Alle Rechte für den Fall der Patenterteilung oder Gebrauchsmustereintragung vorbehalten. © Kendro Laboratory Products GmbH, Germany
4.6
ERROR TABLE
Aside from the current error, outputting the unit's error history may be helpful in cor-
recting a problem.
To output the last 10 errors, press the
i
and the
keys. The most recent error is
displayed in position 1, the oldest in position 10 (refer to the section on error storage in
the operating instructions).
Error table
Code Cause
Fault condition
Communications between
display mP and main board
interrupted.
The display is not receiving display values from
the master processor.
42
Main board NVRAM read er-
ror
Default values were loaded.
43
Main board NVRAM read er-
ror
The mirrored values were loaded.
44
NVRAM defect
Values of the measuring cell are not overwrit-
ten, unit runs using default values
54
Set value error
Error in the calculation of the ser values. The
processor performs a “reset”.
55
I²C bus error
Data transfer to the I
2
bus interrupted.
66
Deviation between tempera-
ture probe PT1000 and LM
75 is too large.
(No longer plausible.)
The validity of the temperature signals is no
longer assured because the permissible devia-
tion between the measured values for:
The incubation mode are > + 2 °C, or;
The decontamination mode are > + 5 °C.
77
CO
2
calculation range ex-
ceeded.
The offset value for the CO
2
adjustment
made by the cal function exceeds the
maximum permissible adjustment range of
+ 10.0 % CO
2
The calculated temperature adjustment
factor exceeds the maximum permissible
adjustment range of 0.8 ... 1.2.
88
auto-start error
The total running time (1080 min.) has passed
without the routine concluding, or the maxi-
mum CO
2
countervoltage adjustment value
has been exceeded.
99
Glass door open or door
switch defective.
The door or door switch have remained in the
"open" state for more than 10 min. (The door
switch contact is closed when the glass door is
open!!)
100
Temperature below set value Actual value < set value - 1.0 °C
101
Temperature above set value Actual value > set value + 1.0 °C
(Sample protection function active.)
Summary of Contents for HERAcell 150
Page 82: ...CO2 Incubator HERAcell 150 Operating Instructions 82 16 Certificate of decontamination...
Page 83: ...CO2 Incubator HERAcell 150 Operating Instructions 83 16 Certificate of decontamination...
Page 84: ...Internet http www kendro com...
Page 85: ......
Page 87: ...Valid 10 2005 50047292 B Gas Cylinder Change Over Unit GM 2 Operating instructions GB...
Page 106: ......
Page 152: ......
Page 153: ......
Page 154: ......
Page 155: ......
Page 156: ......
Page 157: ......
Page 158: ......
Page 159: ......
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 237: ...Version 04 2004 50078754 Accessories for CO2 incubators Illustrations and describtion...
Page 249: ...Footprint HERAcell...
Page 250: ...Footprint HERAcell...