HERAcell
®
, HERAcell
®
150 and HERAcell
®
240, Main PCB Replacement
7
3.5 Configuration Settings
(units without any options, small control box)
The following appear when the device is switched on:
The green lamp incorporated the mains power switch lights up.
8-digit check of both displays, and all LEDs are on.
If High/Low humidity has been configured, the associated LED will also light
up during the 8-digit check.
The version of the current software will be displayed:
P 1 (display PCB):
e.g., 004, that is, Version 4
P 2 (sensor PCB):
e.g., 015, that is, Version 15
P 3 (main PCB):
e.g., 202, that is, Version 202
Pr (parameter set)
e.g.. 004, that is HERAcell, 120V, Cu
The device then displays the current actual values of temperature and CO
2
.
Next, select the appropriate parameter set for your unit from the table below:
Device assignment to the control parameter sets
Parameter set no. Device
Note
1
HERAcell, 230 V, SS
SS = stainless steel
2
HERAcell, 230 V, Cu
3
HERAcell, 120 V, SS
SS = stainless steel
4
HERAcell, 120 V, Cu
5
HERAcell 240, 230 V, SS
SS = stainless steel
6
HERAcell 240, 230 V, Cu
7
HERAcell 240, 120 V, SS
SS = stainless steel
8
HERAcell 240, 120 V, Cu
9
10
11
12
HERAcell 240, 230 V, SS
HERAcell 240, 230 V, Cu
HERAcell 240, 120 V, SS
HERAcell 240, 120 V, Cu
Parameter sets 9 -12, booster
switched off for 6 door screen
Inputting the selected parameter set in function level 601, sublevel 1:
1) The function levels are accessed by simultaneously pressing and holding the
cal
- key,
the
i
- key, and the
auto start
- key for at least 5 sec. The program then jumps to function
level 0.
2) Holding down the
cal
- key and pressing the ? key will take you to function level 601.
Summary of Contents for HERAcell 150
Page 82: ...CO2 Incubator HERAcell 150 Operating Instructions 82 16 Certificate of decontamination...
Page 83: ...CO2 Incubator HERAcell 150 Operating Instructions 83 16 Certificate of decontamination...
Page 84: ...Internet http www kendro com...
Page 85: ......
Page 87: ...Valid 10 2005 50047292 B Gas Cylinder Change Over Unit GM 2 Operating instructions GB...
Page 106: ......
Page 152: ......
Page 153: ......
Page 154: ......
Page 155: ......
Page 156: ......
Page 157: ......
Page 158: ......
Page 159: ......
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 237: ...Version 04 2004 50078754 Accessories for CO2 incubators Illustrations and describtion...
Page 249: ...Footprint HERAcell...
Page 250: ...Footprint HERAcell...