4.7.2 TEMPERATURE MEASUREMENT/REGULATING CIRCUIT – INCUBATION
MODE
Short description:
Sample chamber with controlled, directly mounted heating lines. The unit's air
jacket supports the temperature equilibrium.
Exterior door with mounted heating lines to prevent condensation from forming
on the glass.
P-based temperature regulator with PT1000 as the temperature sensor (inte-
grated in the sensor block).
P-based sample protection function with an upper limit band value that serves
as the trigger threshold to activate the software module.
Plausibility check of the temperature sensor signals (PT1000 and digital sensor,
LM 75, in the sensor block).
The individual heating circuits are activated separately, and for varying lengths
of time.
Circuit diagram:
See circuit diagrams in Chapter 3.
Technical specifications – incubation mode:
Nominal voltage:
230 VAC or 120 VAC
Power consumption:
0.63
kW (HERAcell 240, 230 VAC)
0.64
kW (HERAcell 240, 120 VAC)
Heater resistors
Chapter 6: Metering and Test List
Summary of Contents for HERAcell 150
Page 82: ...CO2 Incubator HERAcell 150 Operating Instructions 82 16 Certificate of decontamination...
Page 83: ...CO2 Incubator HERAcell 150 Operating Instructions 83 16 Certificate of decontamination...
Page 84: ...Internet http www kendro com...
Page 85: ......
Page 87: ...Valid 10 2005 50047292 B Gas Cylinder Change Over Unit GM 2 Operating instructions GB...
Page 106: ......
Page 152: ......
Page 153: ......
Page 154: ......
Page 155: ......
Page 156: ......
Page 157: ......
Page 158: ......
Page 159: ......
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 237: ...Version 04 2004 50078754 Accessories for CO2 incubators Illustrations and describtion...
Page 249: ...Footprint HERAcell...
Page 250: ...Footprint HERAcell...