HERAcell
®
, HERAcell
®
150 and HERAcell
®
240, Main PCB Replacement
6
3. Main PCB Replacement
3.1 Disconnecting the Device from the Mains Power Supply
1.
Switch the device off at the main power switch.
2.
Unplug the mains power cord and secure it to prevent accidental reconnec-
tion.
3.2 Removing the Main PCB
1.
Remove the circuit box cover.
2.
Remove the interface retainer screws.
3.
Remove the Philips head screw on mains inlet connector.
4.
Unplug the orange plug-in connector (X2 – X10) for the mains power circuit
on the PCB.
5.
Disconnect the two door switch leads from terminal block JP2. To do this,
press on the spring terminals with a small screwdriver.
6.
If installed, disconnect the four water level sensor leads from terminal block
JP1. To do this, press on the spring terminals with a small screwdriver.
7.
Unplug the bus cables at the plug-in connections (JP3 / JP4 / JP 15). Press
the lockinng hook on the connectors to release.
8.
Disconnect the two CO
2
solenoid leads from terminal block JP5. To do this,
press on the spring terminal with a small screwdriver.
9.
If installed, unplug the air pump leads at connector JP 11. Press the
lockinng hook on the connectors to release.
10.
Unplug the green/yellow ground wire at the PCB.
11.
Using a Philips head screwdriver, remove the PCB retainer screws (7
screws).
12.
Remove the main PCB from the circuit box.
3.3 Installing the New PCB
1.
Install the new PCB in the reverse order to the removal instructions given in
Section 3.2.
2.
An electrical safety check must be performed after electrical components
have been removed / installed.
3.4 Restarting the Device
1.
Plug the mains power cord into the mains inlet connector.
2.
Switch the device on at the mains power switch.
3.
Resume normal operations.
Summary of Contents for HERAcell 150
Page 82: ...CO2 Incubator HERAcell 150 Operating Instructions 82 16 Certificate of decontamination...
Page 83: ...CO2 Incubator HERAcell 150 Operating Instructions 83 16 Certificate of decontamination...
Page 84: ...Internet http www kendro com...
Page 85: ......
Page 87: ...Valid 10 2005 50047292 B Gas Cylinder Change Over Unit GM 2 Operating instructions GB...
Page 106: ......
Page 152: ......
Page 153: ......
Page 154: ......
Page 155: ......
Page 156: ......
Page 157: ......
Page 158: ......
Page 159: ......
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 237: ...Version 04 2004 50078754 Accessories for CO2 incubators Illustrations and describtion...
Page 249: ...Footprint HERAcell...
Page 250: ...Footprint HERAcell...