Gas Tight Screen for the HERAcell
®
and HERAcell
®
150, Retrofit
Version: 07.2003 / 50078424
12
3.6 Configuration control- and rule system (softwareversion 210)
Beside the installation of the gas tight
screen you have to configure the
appropiate set of parameters.
Enter
funktion level 21
and choose
the right set of the parameters.
Information:
After switching on the incubator,
tha actual set of the parameters is
indicated during the start routine.
Upper display:
Pr
Lower display:
Parameter
General description
Input
1. To access the function levels
simultaneously press the
cal
– key, the
i
– key, and the
auto-start
– key, and
hold them down for at least 5 seconds.
The programm enters function level 0.
2. To move to the various function levels,
press and hold down the
cal
– key and
use the
- key.
Go to function level 21.
3. To access a particular sublevel, press
and release the
cal
– key until you
rech the desired sublevel.
1.
2.
3.
Enter sublevel 2.
See chart 1 function level 21.
Press the
cal
and
- key, set
value 1.
(Set of parameters unlocked)
Enter sublevel 4.
Press the
cal
and
- key, set
value 1.
Enter sublevel 10.
Press the
cal
and
- key, set
value 1. (writing EPROM)
4. After changing the set of parameters
the unit restarts. If nothing has been
changed continue with 4.1.
4.1 To exit the function level:
- Press the
C
– key or the
.
% CO
2
– key or
– wait 5 seconds without pressing
. a key.
Summary of Contents for HERAcell 150
Page 82: ...CO2 Incubator HERAcell 150 Operating Instructions 82 16 Certificate of decontamination...
Page 83: ...CO2 Incubator HERAcell 150 Operating Instructions 83 16 Certificate of decontamination...
Page 84: ...Internet http www kendro com...
Page 85: ......
Page 87: ...Valid 10 2005 50047292 B Gas Cylinder Change Over Unit GM 2 Operating instructions GB...
Page 106: ......
Page 152: ......
Page 153: ......
Page 154: ......
Page 155: ......
Page 156: ......
Page 157: ......
Page 158: ......
Page 159: ......
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 237: ...Version 04 2004 50078754 Accessories for CO2 incubators Illustrations and describtion...
Page 249: ...Footprint HERAcell...
Page 250: ...Footprint HERAcell...