BETRIEBSANLEITUNG
HERAline
Seite 13 von 14
Datum: 13.04.04
Bearbeiter / Abteilung:
Identnummer:
interner Vermerk
Detlef Dornseiff / DT
50 076 300
50076300 Index
A.doc
11.
Description for HERAcell 150 and HERAcell 240 (O2 – models)
The HERAline analog interface converts the digital values of temperature and CO2
from the RS 232 interface of the incubator in 4 analog signals, having 4....20 mA, by
using a 4 channel D/A converter together with a microprocessor. The resolution of
the D/A conversation is 10 bit for each channel, which means totally 1024 steps.
Since the RS 232 will be used for the HERAline interface, it will not be possible to
connect an additional computer to the RS 232 interface of the incubator.
The analog outputs are connected to a female 15 pole D-Sub connector, which is
located on the analog interface. External devices can be connected to the interface
by using the standart RGB/S video cable which commands with the HERAline
interface.
The analog output signals are:
Red
„R“ Actual temperature
incubation mode
Actual temperature
desinfection mode
0,16 mA/°C
0,16 mA/°C
0...55 °C
0...90°C
=
=
4...12,8 mA
4...18,4 mA
Green
„G“ Actual CO2 –
concentration
0,8 mA/%
0...20% CO2 = 4...20 mA
Blue
„B“ Actual O2 –
concentra
0,16 mA/%
0,16 mA/%
1...21 % O2
5...90 % O2
=
=
4,16...7,36 mA
4,8...18,4 mA
Black
„S“ Not connected
It will be possible to connect the external devices directly to a male 15 pole D-Sub
connector which fits to the female output connector of the incubator.
The outputs are connected to:
Actual temperature Pin 1
(+)
Actual CO2
Pin 2
(+)
Set temperature
Pin 3
(+)
Set CO2
Pin 13
(+)
Ground (-)
Pin 4,6,7,10 und 11
(-)
Summary of Contents for HERAcell 150
Page 82: ...CO2 Incubator HERAcell 150 Operating Instructions 82 16 Certificate of decontamination...
Page 83: ...CO2 Incubator HERAcell 150 Operating Instructions 83 16 Certificate of decontamination...
Page 84: ...Internet http www kendro com...
Page 85: ......
Page 87: ...Valid 10 2005 50047292 B Gas Cylinder Change Over Unit GM 2 Operating instructions GB...
Page 106: ......
Page 152: ......
Page 153: ......
Page 154: ......
Page 155: ......
Page 156: ......
Page 157: ......
Page 158: ......
Page 159: ......
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 237: ...Version 04 2004 50078754 Accessories for CO2 incubators Illustrations and describtion...
Page 249: ...Footprint HERAcell...
Page 250: ...Footprint HERAcell...