4.7.8 O
2
METERING AND REGULATION CIRCUIT
Short description:
Zink oxide sensor located in the sample chamber and with replaceable, plugged-in
sensor head.
auto–cal routine for automatic adjustment is integrated into the auto-start routine.
µP-based O
2
regulator.
O
2
or N
2
input via the same connection spout.
After entering the sample chamber, the gas is dispersed across the surface of the
water to increase the rate of humidity regeneration.
The O
2
or N
2
gas input is metered by a valve. The gas flowrate is also restricted by
a capillary opening in the connection spout.
Prior to entering the unit, the O
2
or N
2
passes through a sterile filter.
The O
2
metering and regulation circuit can be manually switched off by adjusting
the set value to < 5 % (for the O
2
range of 5 ... 90 %) or adjusting the set value >21
% (for the O
2
range of 1 ... 21 %).
The initial startup ‘‘run‘‘ phase of the O
2
sensor is approx. 5 min. During this time,
CO
2
gas addition is suppressed.
Circuit diagram:
See circuit diagrams in Chapter 3.
Technical specifications:
Gas flowrate:
O
2
~ 11 l/min
N
2
~ 13 l/min
Gas input pressure:
1 bar
Summary of Contents for HERAcell 150
Page 82: ...CO2 Incubator HERAcell 150 Operating Instructions 82 16 Certificate of decontamination...
Page 83: ...CO2 Incubator HERAcell 150 Operating Instructions 83 16 Certificate of decontamination...
Page 84: ...Internet http www kendro com...
Page 85: ......
Page 87: ...Valid 10 2005 50047292 B Gas Cylinder Change Over Unit GM 2 Operating instructions GB...
Page 106: ......
Page 152: ......
Page 153: ......
Page 154: ......
Page 155: ......
Page 156: ......
Page 157: ......
Page 158: ......
Page 159: ......
Page 160: ......
Page 161: ......
Page 162: ......
Page 163: ......
Page 164: ......
Page 165: ......
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......
Page 171: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 237: ...Version 04 2004 50078754 Accessories for CO2 incubators Illustrations and describtion...
Page 249: ...Footprint HERAcell...
Page 250: ...Footprint HERAcell...