12.6. Pumps
Using this tab, the wash pump and waste valve can
be controlled.
12.6.1. Wash pump
There are two buttons:
•
Start pump wash, starts the pump for the
configured time in "Pump time (ms)".
•
Stop pump wash, stops the pump.
12.6.2. Waste Valve
The waste valve can be used to empty the filter from the pressure regulator. Hold the button for approximately
5 seconds to empty the filter.
12.7. Pipetting arm channel movements
Using this visualization, all the different Anybus components can be controlled.
There are 4 tabs:
•
Main (see Chapter 12.2 Main)
•
General
•
Channel 1-4
12.7.1. General
Using this tab, the different parts on the Anybus can
be controlled and checked directly using control and
status words.
12.7.2. Control word
Using the control word, different parts on the Anybus
network can be controlled.
Byte 1 is the control word.
Byte 2 is the "NodeID", which is the address where the control word should be send to:
•
No nodes selected
•
Channel 1
•
Channel 2
•
Channel 3
•
Channel 4
•
Anybus
•
All Nodes (default)
Overview control words:
•
Toggle control word from 0-128 - Set halt state Anybus
•
Toggle control word from 0-129 - Set Anybus active
12.7.3. Status word
Show that the status word of the selected NodeID, where Byte 1 is the status of the selected NodeID and
Byte 2 is the error byte, indicating which Node has an error.