Timeout triggering
plate barcode
scanner
Error
• Error is added to Alarm Manager "E401: Timeout triggering plate
barcode scanner".
• Acknowledge alarm in the alarm manager.
• Initialize the system before Autosampler III can be restarted.
• If problem persists, contact support.
No connection
with PLC
Error
• Error is added to Alarm Manager "OPC-UA Connection failed".
• Restart HMI application.
• If problem persists, contact support.
No connection
with shaker
Error
• Error is added to Alarm Manager "Could not connect to Shaker".
• Alarm needs to be acknowledged before Autosampler III can
be restarted.
• If problem persists, contact support.
• Autosampler III can still be run using the method for running
without shaker.
Shaker not initialized Error
• Error is added to Alarm Manager "Shaker could not initialize".
• Alarm needs to be acknowledged before Autosampler III can
be restarted.
• If problem persists, contact support.
• Autosampler III can still be run using the method for running
without shaker.
No connection with
pipet pumps
Error
• Error is added to Alarm Manager "Could not connect with pipetting
pumps".
• Alarm needs to be acknowledged before Autosampler III can be
restarted.
• If problem persists, contact support.
No connection with
reagent pumps
Error
• Error is added to Alarm Manager "Could not connect with
reagent pumps".
• Acknowledge alarm in the alarm manager.
• Initialize the system before Autosampler III can be restarted.
• If problem persists, contact support.
No connection with
camera
Error
• Error is added to Alarm Manager "Connection to Cognex
camera failed".
• Alarm needs to be acknowledged before Autosampler III can
be restarted.
• If problem persists, contact support.
No connection with
pipetting channels
Error
• Error is added to Alarm Manager "Could not write to OPC-UA".
• Alarm needs to be acknowledged before Autosampler III can
be restarted.
• If problem persists, contact support.
Air pressure low / no
air pressure
Error
• Error is added to Alarm Manager "E003: Air pressure not available".
• Alarm needs to be acknowledged before Autosampler III can
be restarted.
• Wait for a moment, maybe the soft start wasn't ready yet.
• If not, check air.
• If problem persists, contact support.
Error with one of the
pipetting pumps
Error
• Error is added to Alarm Manager "Error Pipetting pump
'Pump Address' 'serial Response'".
• Acknowledge alarm in the alarm manager.
• Initialize the system before Autosampler III can be restarted.
• If problem exists, contact support.