5.15. Conditions of use
The products to be processed must be undamaged.
Environmental conditions:
•
The Autosampler III must be set up in a clean and dry environment.
In addition, the following requirements must be met:
–
Minimum ambient temperature: 0º C
–
Maximum ambient temperature: 40º C
–
Maximum relative humidity: 80%
Electrical energy requirements:
•
Rated voltage: 400 V
•
Power connection: 11 kVA
•
Rated frequency: 50 Hz
•
For further electrotechnical data see the electrical diagrams.
Requirements related to pneumatic installation.
•
Rated working pressure: 7 bar.
•
Compressed air quality according to DIN ISO 8573-1
5.16. Description of the operator's workspace
The user's workplace is indicated by the following
picture. The workplace is visible in the image to
the right.
5.17. Personal protective equipment
When using the Autosampler III, samples/recipients
are used in which biological and chemical material
is stored. It is up to the user to apply personal
protective equipment as necessary in relation to
these samples. It is advised to always wear a lab
coat, gloves and safety glasses when working with
the Autosampler III.
5.18. Restrictions on the use of the Autosampler III
The Autosampler III can be used for pipetting sample material from specified packaging to a shallow well
plate. Liquids to be pipetted must be water-based and must not contain flammable substances.
Indication of a good working environment.