www.hygiena.com
93
Error with one of the
reagent pumps
Error
• Error is added to Alarm Manager "Error reagent pump
'Pump Address' 'serial Response'".
• Acknowledge alarm in the alarm manager.
• Initialize the system before Autosampler III can be restarted.
• If problem exists, contact support.
Reagent 1 empty
warning
Warning
• Warning is added to Alarm Manager "Warning: Reagent volume 1,
less than 'volume' mL available".
• Autosampler III will continue.
Reagent 1 empty
error
Non-fatal
error
• Non-fatal error is added to Alarm Manager "Error: Reagent volume 1,
less than 'volume' mL available".
• Autosampler III is paused.
• Fill reagent 1.
• Alarm needs to be acknowledged.
• Press "Start" to continue run.
Reagent 2 empty
warning
Warning
• Warning is added to Alarm Manager "Warning: Reagent volume 2,
less than 'volume' mL available".
• Autosampler III will continue.
Reagent 2 empty
error
Non-fatal
error
• Non-fatal error is added to Alarm Manager "Error: Reagent volume 2,
less than 'volume' mL available".
• Autosampler III is paused.
• Fill reagent 2.
• Alarm needs to be acknowledged.
• Press "Start" to continue run.
Error while adding
samples to file
Non-fatal
error
• Make sure the file is not opened by another application.
Needles broken
Failure
(to be
observed
only)
• Replace needles as described in this manual.
Tubing leaking
Failure
(to be
observed
only)
• Check connection, clean or replace as described in this manual.
Syringe leaking
Failure
(to be
observed
only)
• Inspect the pump and clean or replace as described in this manual.