Site requirements
Vivid S70 / S60
–
User Manual
3-3
BC092760-1EN
01
Operating Environment
Ensure that there is sufficient air flow around the ultrasound
system when installed or operated.
Environmental requirements
The ultrasound system requires constant maintenance of its
operational environment. Different temperature and humidity
requirements are specified for operation, storage and
transportation.
NOTE:
System may be operated at an altitude of up to 3000 meters.
Electromagnetic interferences
The ultrasound system is approved for use in hospitals, clinics,
and other environmentally qualified facilities, in terms of the
prevention of radio wave interference. Operation of the system
in an inappropriate environment can cause electronic
interference to radios and television sets situated near the
medical equipment.
Ensure that the system is protected from electromagnetic
interferences as follows:
•
Operate the system at least 4.5 meters (15 feet) away from
equipment that emits strong electromagnetic radiation.
•
Shield the system when operating it in the vicinity of radio
broadcasting equipment, if necessary.
Requirement
Temperature
Humidity
Air Pressure
Operational
10–35
°
C
30–85%
700–1060 hPa
Storage
-20–60
°
C
10–95%
700–1060 hPa
Transport
-20–60
°
C
10–95%
700–1060 hPa
Summary of Contents for Vivid S70
Page 8: ...i 6 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 62: ...Safety 2 30 Vivid S70 S60 User Manual BC092760 1EN 01 For China only...
Page 63: ...Device labels Vivid S70 S60 User Manual 2 31 BC092760 1EN 01...
Page 64: ...Safety 2 32 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 349: ...4D Auto MVQ Vivid S70 S60 User Manual 8 71 BC092760 1EN 01 Figure 8 37 Set Landmark stage...
Page 382: ...Measurements and Analysis 8 104 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 670: ...Probes 13 54 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 704: ...Maintenance 15 14 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 708: ...Index 4 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 709: ......