Contrast Imaging
Vivid S70 / S60
–
User Manual
5-45
BC092760-1EN
01
For further information on quantification, see ‘Quantitative
Analysis’ on
Contrast Echocardiography Box
Contrast Echocardiography Box
: the European Society of
Cardiology provides useful information on contrast
echocardiography on the Internet here: https://
www.escardio.org/Guidelines-&-Education/Practice-tools/
EACVI-toolboxes/Contrast-Echo/
Contrast-Echocardiography-Box
WARNING
Misdiagnosis based on image artifacts
Misdiagnosis in ultrasound contrast images may be caused by
several artifacts, most importantly:
Motion artifacts
: gives rise to signals independently of
contrast presence. This may be caused by patient movement;
including respiration, or by probe movement influenced by the
operator.
Regional drop outs
: caused by unintentional destruction of
the contrast agent, too low concentration of contrast agent,
poor acoustic penetration due to rib/lung shadows or system
failing to detect the contrast agent due to erroneous settings
induced by the operator.
Tissue harmonics
: gives contrast-like signals independently
of the presence of contrast agent.
Summary of Contents for Vivid S70
Page 8: ...i 6 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 62: ...Safety 2 30 Vivid S70 S60 User Manual BC092760 1EN 01 For China only...
Page 63: ...Device labels Vivid S70 S60 User Manual 2 31 BC092760 1EN 01...
Page 64: ...Safety 2 32 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 349: ...4D Auto MVQ Vivid S70 S60 User Manual 8 71 BC092760 1EN 01 Figure 8 37 Set Landmark stage...
Page 382: ...Measurements and Analysis 8 104 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 670: ...Probes 13 54 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 704: ...Maintenance 15 14 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 708: ...Index 4 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 709: ......