Image Optimization
5-42
Vivid S70 / S60
–
User Manual
BC092760-1EN
01
Figure 5-22. The LV Contrast Touch panel
Using LV Contrast
The LV Contrast application works with the M5Sc-D, 3Sc-RS,
6VT-D and 6Tc-RS probes.
1. Press
Probe
on the Touch panel.
A list of the connected probes with respective applications is
displayed.
2. Select
LV Contrast
application.
3. Turn on and adjust
Tissue Visualization
if needed, to
re-orient the probe after the application has been selected.
4. Once oriented, turn off
Tissue Visualization
for optimized
contrast detection and perform the acquisition.
NOTE:
If the amount of contrast injected is too high, you may press
Flash
to destroy some of the contrast.
5. To continuously monitor the view orientation while acquiring
contrast data, press
Simultan.
to activate a side-by-side
view of tissue and contrast visualization.
WARNING
Always read and follow carefully the manufacturer instructions
on the contrast agent label.
Summary of Contents for Vivid S70
Page 8: ...i 6 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 62: ...Safety 2 30 Vivid S70 S60 User Manual BC092760 1EN 01 For China only...
Page 63: ...Device labels Vivid S70 S60 User Manual 2 31 BC092760 1EN 01...
Page 64: ...Safety 2 32 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 349: ...4D Auto MVQ Vivid S70 S60 User Manual 8 71 BC092760 1EN 01 Figure 8 37 Set Landmark stage...
Page 382: ...Measurements and Analysis 8 104 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 670: ...Probes 13 54 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 704: ...Maintenance 15 14 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 708: ...Index 4 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 709: ......