Connectivity configuration
Vivid S70 / S60
–
User Manual
12-53
BC092760-1EN
01
From DICOM images the user can select between:
•
No extra info
•
Add visible patient info in the DICOM images
: displays
patient information (name, date of birth and ID) on DICOM
images.
•
Add title bar
: adds the Title bar to the DICOM images. Must
be enabled in order for images from the 4D Auto AVQ tool to
contain traces, bullseye, 3D model and measurements as
part of the stored DICOM Multiframe/Singleframe files when
using a dataflow with RAW data disabled.
When
Use high DICOM resolution
is enabled, DICOM images
are stored using a higher pixel density. Use this setting when
exporting to systems accessed by high definition DICOM
viewing stations.
NOTE:
Using high DICOM resolution will double the file size of the
DICOM data when using standard compression settings. Such
files will consume more disk space and also slow down storage,
recall and transfer of files.
Use
Movie optimized
to enable temporal filtering on DICOM
images for enhanced image representation on DICOM
workstations.
Color management
provides a selection of gamma settings for
optimized representation of the Vivid images on DICOM
workstations.
Summary of Contents for Vivid S70
Page 8: ...i 6 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 62: ...Safety 2 30 Vivid S70 S60 User Manual BC092760 1EN 01 For China only...
Page 63: ...Device labels Vivid S70 S60 User Manual 2 31 BC092760 1EN 01...
Page 64: ...Safety 2 32 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 349: ...4D Auto MVQ Vivid S70 S60 User Manual 8 71 BC092760 1EN 01 Figure 8 37 Set Landmark stage...
Page 382: ...Measurements and Analysis 8 104 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 670: ...Probes 13 54 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 704: ...Maintenance 15 14 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 708: ...Index 4 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 709: ......