Measurements and Analysis
8-66
Vivid S70 / S60
–
User Manual
BC092760-1EN
01
By creating several control points the sampling path can be
bent to follow the myocardium.
8. Press
Select
twice to end the sampling path.
9. Create a sampling path in scan plane 2 and 3.
A TSI color coded surface model is displayed in the
Geometric model
.
10. To create a dynamic model, scroll to another frame in the
same heart cycle and create a sampling path in each scan
plane.
11. Press
2D Freeze
to run the model.
To edit the sampling path
1. Place the cursor over a control point.
2. Press
Select
twice (double-click) and move the control point
to a new position using the trackball.
3. Press
Select
to place the control point to its new position.