4D/Multi-plane LV
Vivid S70 / S60
–
User Manual
8-59
BC092760-1EN
01
Figure 8-30. The Volume waveform screen
Approval
1. Press
Approve & Exit
to store the measurements shown in
the
Measurement result
table.
The measurements are transferred to the Worksheet.
To exit without approving, press
Cancel
.
NOTE:
Measurements that are not approved will not be saved.
Calculation formula and measurement accuracy
Refer to the Reference manual for calculation formula
information and ‘Measurement accuracy’ on
for
measurement accuracy information.
CAUTION
The measurement should not be approved if a significant part
(more than 25%) of the left ventricular wall or the detected
contours are outside the ultrasound sector.