Stress Echo
7-6
Vivid S70 / S60
–
User Manual
BC092760-1EN
01
Parameters
window on the right side of the screen (see
).
Figure 7-3. Display of the Reference loop during acquisition
Protocol Pause function
During the stress acquisition it is possible to temporarily exit the
protocol acquisition mode to acquire images in any mode
outside the stress protocol.
1. To temporarily exit the protocol mode, press
Protocol
Pause
on the Touch panel.
2. Acquire the desired images outside the protocol.
3. Press
Protocol Pause
to restart the protocol acquisition
mode and resume the stress acquisition.
Selecting a view during acquisition
A fixed protocol is provided for scanning, based on the selected
template. The system automatically highlights the next view to
be acquired in the template matrix, as images are stored.
However, the order of scanning may be changed manually as
follows.
1. Current acquisition loop
2. Corresponding reference loop
Summary of Contents for Vivid S70
Page 8: ...i 6 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 62: ...Safety 2 30 Vivid S70 S60 User Manual BC092760 1EN 01 For China only...
Page 63: ...Device labels Vivid S70 S60 User Manual 2 31 BC092760 1EN 01...
Page 64: ...Safety 2 32 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 349: ...4D Auto MVQ Vivid S70 S60 User Manual 8 71 BC092760 1EN 01 Figure 8 37 Set Landmark stage...
Page 382: ...Measurements and Analysis 8 104 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 670: ...Probes 13 54 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 704: ...Maintenance 15 14 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 708: ...Index 4 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 709: ......