Image Optimization
5-38
Vivid S70 / S60
–
User Manual
BC092760-1EN
01
AVC + 150 ms, AVC + 200 ms, MVO, MVO + 100 ms,
MVO + 160 ms, MVO + 200 ms, MVO + 260 ms,
ES - 200 ms, ES - 150 ms, ES - 100 ms, ES - 50 ms,
ES + 50 ms, ES + 100 ms, ES + 150 ms, ES + 200 ms or
Manual control.
NOTE:
Manual adjustment of TSI start and TSI end markers is
available in Q Analysis. To store the modified marker
settings, press
Store
and choose the configuration setting
Manual control
to avoid automatic adjustment of the
markers.
To configure TSI markers:
1. Press
Utility
/
Config
on the Touch panel and select the
category
Meas/Text
.
2. In the
Measure category
, select the sheet
Advanced
.
3. In the
Application specific parameters section
adjust TSI
start and TSI end parameters by selecting a new value from
the combo menu displayed upon selection.