Connectivity configuration
Vivid S70 / S60
–
User Manual
12-49
BC092760-1EN
01
Refer to the Vivid S70 / S60 Reference manual for a
complete list of supported parameters.
•
The following modes: 2D, M-mode, Color Flow, PW Doppler,
CW Doppler, 3D and TDI.
•
Not Modified Simpson method or Bullet methods.
Refer to the Vivid S70 / S60 Reference manual for a
complete list of supported methods.
•
Basic derivations (Average, Last, Min and Max), no
references between the derived measurements and the
ones they were made from.
•
Wall Motion Scoring: individual segment scores only
according to 16-segment model, no graded Hypokinesis
(only Hypokinesis is used).
DICOM SR must be activated for each DICOM device.
1. Press
Utility
/
Config
on the control panel and log on as
administrator.
2. Select the
Connectivity
category and
Dataflow
subgroup.
The
Dataflow
sheet is displayed (see Figure 12-29).
3. Select the DICOM dataflow to configure in the
Dataflow
pull-down menu.
4. Select a DICOM storage device and press
Properties
.
The
Properties
window for the selected DICOM storage
device is displayed.
Figure 12-34. DICOM SR Settings
5. Check the option
Allow SR
to enable DICOM SR.
The following additional options are available:
•
Allow SR private data
: include current exam data in a
private format within DICOM SR to retain
measurements and complete exam information when
recalling an exam from a DICOM environment. This is
especially important if EchoPAC Plug-in is used in the
Summary of Contents for Vivid S70
Page 8: ...i 6 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 62: ...Safety 2 30 Vivid S70 S60 User Manual BC092760 1EN 01 For China only...
Page 63: ...Device labels Vivid S70 S60 User Manual 2 31 BC092760 1EN 01...
Page 64: ...Safety 2 32 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 349: ...4D Auto MVQ Vivid S70 S60 User Manual 8 71 BC092760 1EN 01 Figure 8 37 Set Landmark stage...
Page 382: ...Measurements and Analysis 8 104 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 670: ...Probes 13 54 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 704: ...Maintenance 15 14 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 708: ...Index 4 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 709: ......