Stress Echo
7-24
Vivid S70 / S60
–
User Manual
BC092760-1EN
01
Figure 7-11. Vpeak QTVI Stress display
Turn-off the Vpeak measurement tool
1. Select the
V
button in the peak view scoring diagrams.
V-peak measurement interpretation
The systolic Vpeak in the tissue velocity profile is automatically
detected and highlighted by a vertical bar (Figure 7-11). The
automatically detected Vpeak should be visually verified by the
user. In addition Vpeak thresholds are displayed as color-coded
horizontal lines (Figure 7-11). These thresholds represent
statistical guideline values for peak velocity at peak stress level
(Dobutamine stress procedure) for the three apical views. Only
threshold values for basal and mid-segments for each apical
view are defined (see reference 1 on page 7-25). The result is
highlighted by a color-coding of the thresholds lines, the
1. Tissue velocity profile
2. Sampling point
3. Current segment
4. Vpeak threshold for current segment
5. Color-coded tissue velocity:
Color-coding (velocity thresholds and tissue):
•
Green
: velocities above threshold value
•
Yellow
: velocities near threshold (0 to -10% interval)
•
White
: velocities below threshold value - 10%
Summary of Contents for Vivid S70
Page 8: ...i 6 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 62: ...Safety 2 30 Vivid S70 S60 User Manual BC092760 1EN 01 For China only...
Page 63: ...Device labels Vivid S70 S60 User Manual 2 31 BC092760 1EN 01...
Page 64: ...Safety 2 32 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 349: ...4D Auto MVQ Vivid S70 S60 User Manual 8 71 BC092760 1EN 01 Figure 8 37 Set Landmark stage...
Page 382: ...Measurements and Analysis 8 104 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 670: ...Probes 13 54 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 704: ...Maintenance 15 14 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 708: ...Index 4 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 709: ......