Probes
13-22
Vivid S70 / S60
–
User Manual
BC092760-1EN
01
Figure 13-6. Cleaning the probe using a brush
Figure 13-7. Probe Lens Examples
5. Continue brushing the probe for not less than the minimum
contact time listed on the detergent manufacturer’s label.
6. Visually inspect the probe for soil. Repeat Steps 3 through 5
until all visible soil has been removed from the surface of the
probe.
7. Rinse the probe under running warm potable water (30 -
40°C) for not less than 2 minutes. Scrub the surface of the
probe with a clean, soft, nylon bristle brush from the base of
the cable strain relief to the distal tip.
CAUTION
Do not use the brush on the probe lens.
CAUTION
Do not use the brush on the probe lens.