derived from endocardial deformation are suffixed Endo,
such as for instance GPeakSysSL(Avg)_Endo
•
Myocardial/full wall deformation is assessed by performing
analysis of the whole ROI trace. Measurements derived
from myocardial deformation are suffixed Full, such as for
instance GPeakSysSL(Avg)_Full
It is possible to change the reference layer preference using the
Config screen
NOTE:
If attempting to reprocess an analysis using a different reference
layer a warning will be shown informing that the values in the
analysis will change.
AFI configuration
It is possible to configure some of the AFI controls to modify the
workflow slightly.
1. Press
Utility/Config
on the Touch panel, select the category
Meas/Text
and the
Advanced Quantification
subgroup.
Select the
AFI
tab. The following parameters may be
adjusted:
•
Autoprocessing timeout
- The time the operator has to
keep trackball still before automatic start of tracking.
•
BE Mode
– Colormaps that shall be available for the
Bulls Eye.
•
BE Segment model
- Select the preferred Bulls Eye
standard (ASE/AFI, 17/18 segment models).
•
ROI Method
– Select whether fully automatic or 3-Click
ROI shall be the tool default.
•
YoYo
– When adjusting the ROI, select whether a
limited number of frames around the selected ROI
frame should be looped back and forth to ease ROI
adjustment.
•
Strain reference layer
– Select whether the strain values
shall be calculated based on endocardial or full wall
tracking.
•
AVC Stage mode
– Select whether to always open the
AVC Selection stage. If "Auto" is selected, upon analysis
of the APLAX view, the user will be taken to Results
stage after the ROI stage. "Auto" AVC or Event timing
will be used by default. It will still be possible for the user
to re-visit the AVC Selection stage to edit the chosen
AVC.
Summary of Contents for Vivid S70
Page 8: ...i 6 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 62: ...Safety 2 30 Vivid S70 S60 User Manual BC092760 1EN 01 For China only...
Page 63: ...Device labels Vivid S70 S60 User Manual 2 31 BC092760 1EN 01...
Page 64: ...Safety 2 32 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 349: ...4D Auto MVQ Vivid S70 S60 User Manual 8 71 BC092760 1EN 01 Figure 8 37 Set Landmark stage...
Page 382: ...Measurements and Analysis 8 104 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 670: ...Probes 13 54 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 704: ...Maintenance 15 14 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 708: ...Index 4 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 709: ......