Probes
13-42
Vivid S70 / S60
–
User Manual
BC092760-1EN
01
Coupling gels
In order to ensure optimal transmission of energy between the
patient and probe, a conductive gel or couplant must be applied
liberally to the patient where scanning will be performed.
Refer to the Probe Care Card enclosed in the probe case or to
the Internet link below for the latest list of compatible coupling
gels, cleaners and disinfectans.
Precautions: coupling gels should not contain the following
ingredients as they are known to cause probe damage:
•
Methanol, ethanol, isopropanol, or any other alcohol-based
product
•
Mineral oil
•
Iodine
•
Lotions
•
Lanoline
•
Aloe Vera
•
Olive Oil
•
Methyl or Ethyl Parabens (para hydroxybenzoic acid)
•
Dimethylsilicone
•
Polyether glycol based
•
Petroleum
CAUTION
Do not apply gel to the eyes. If there is gel contact to the eye,
flush eye thoroughly with water.
CAUTION
Only use GE-recommended gels. Use of unapproved gels may
result in damage to the probe and void the warranty.
http://www3.gehealthcare.com/en/Products/Categories/Ultrasound/Ultrasound_Probes
Summary of Contents for Vivid S70
Page 8: ...i 6 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 62: ...Safety 2 30 Vivid S70 S60 User Manual BC092760 1EN 01 For China only...
Page 63: ...Device labels Vivid S70 S60 User Manual 2 31 BC092760 1EN 01...
Page 64: ...Safety 2 32 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 349: ...4D Auto MVQ Vivid S70 S60 User Manual 8 71 BC092760 1EN 01 Figure 8 37 Set Landmark stage...
Page 382: ...Measurements and Analysis 8 104 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 670: ...Probes 13 54 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 704: ...Maintenance 15 14 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 708: ...Index 4 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 709: ......