Care and Maintenance
Vivid S70 / S60
–
User Manual
13-23
BC092760-1EN
01
8. Visually inspect the device in a well-lit area to ensure all
surfaces are free from residual cleaning solution. Repeat
Step 7 if visible cleaning solution is observed.
9. Thoroughly dry the probe using a clean lint-free soft and dry
cloth or wipe.
Cable and Connector Manual Cleaning
1. The cable and connector surfaces can be cleaned with the
cleaners or wipes listed in the Probe Care Card.
NOTE:
Use of wipes listed in the Probe Care Card may result in
discoloration of the cable.
2. Wipe the cable with a lint-free cloth dampened with potable
water to remove chemical residue.
Figure 13-8. Cleaning the Probe Cable
WARNING
Use caution when cleaning the connector. This cable connector
should only be cleaned with a slightly dampened cloth or wipe.
Exposure to excessive moisture will result in damage to the
probe and possibly the ultrasound console. DO NOT wet the
connector/console interface surface or labels.
Summary of Contents for Vivid S70
Page 8: ...i 6 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 62: ...Safety 2 30 Vivid S70 S60 User Manual BC092760 1EN 01 For China only...
Page 63: ...Device labels Vivid S70 S60 User Manual 2 31 BC092760 1EN 01...
Page 64: ...Safety 2 32 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 349: ...4D Auto MVQ Vivid S70 S60 User Manual 8 71 BC092760 1EN 01 Figure 8 37 Set Landmark stage...
Page 382: ...Measurements and Analysis 8 104 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 670: ...Probes 13 54 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 704: ...Maintenance 15 14 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 708: ...Index 4 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 709: ......