Image Optimization
5-52
Vivid S70 / S60
–
User Manual
BC092760-1EN
01
UD Speckle reduce
2D (non-Cardiac)
Reduces the unwanted effects of speckle in the ultrasound image. Image speckle
usually appears as a grainy texture in otherwise uniform areas of tissue. Its
appearance is related to image system characteristics, rather than tissue
characteristics, so that changes in system settings, such as probe type, frequency,
depth, and others, can change the appearance of the speckle. Too much speckle
can impair image quality and make it difficult to see the desired detail in the image.
Likewise, too much filtering of speckle can mask or obscure desired image detail.
Extra care must be taken to select the optimal Speckle reduction level.
Virtual Apex
2D (Cardiac)
Improves near field imaging, allowing increased visibility up to the width of the full
probe aperture close to the surface.
Virtual Convex
2D, CF, Doppler
Virtual Convex extends the field of view for linear probes. Enable/disable
independently for each mode.
Compound
2D
Combines three or five frames from different steering angles into a single frame,
reducing speckle noise and clutter, and providing continuity of specular reflectors.
Edge Enhance
2D, M-Mode
(non-cardiac)
Controls image processing related to the extent of edge enhancement applied to
an image.
PRF
2D
Affects the level of reverberations in the image. When turned on, the frame rate (or
the number of focal zones) will decrease, while the reverberations will be
attenuated.
Power
2D, M-Mode, CF,
Doppler, TVI, TT,
SRI, SI, TSI
Controls the amount of acoustic power applied in all modes. When power is set to
maximum, it is equal to or less than the maximum acoustic power permitted by the
FDA. The Thermal Index (TI) and the Mechanical Index (MI) are displayed on the
screen.
When power is reduced, it reduces the signal-to-noise ratio, so that the image may
become noisier.
Horizontal sweep
M-Mode, Doppler
Adjusts the horizontal refresh rate of the M-Mode or Doppler area of the display.
Horizontal sweep is available in live and cine replay.
Summary of Contents for Vivid S70
Page 8: ...i 6 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 62: ...Safety 2 30 Vivid S70 S60 User Manual BC092760 1EN 01 For China only...
Page 63: ...Device labels Vivid S70 S60 User Manual 2 31 BC092760 1EN 01...
Page 64: ...Safety 2 32 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 349: ...4D Auto MVQ Vivid S70 S60 User Manual 8 71 BC092760 1EN 01 Figure 8 37 Set Landmark stage...
Page 382: ...Measurements and Analysis 8 104 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 670: ...Probes 13 54 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 704: ...Maintenance 15 14 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 708: ...Index 4 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 709: ......