4D-Mode
Vivid S70 / S60
–
User Manual
6-9
BC092760-1EN
01
NOTE:
4D acquisition may also be done in Multi-Slice mode
(‘Multi-Slice acquisition’ on page 6-10) and 4D Zoom
prepare mode (‘4D Zoom prepare acquisition’ on
page 6-13).
The volume acquisition is started.
4. The following parameters may be adjusted during
acquisition:
•
Volume size
: adjust the volume size. An increase of the
volume size will reduce the volume rate.
•
Volume shape
: adjust the ratio between the elevation
and azimuth widths. Changing the volume shape from
default will increase the volume rate.
•
Elevation tilt
: tilt the volume in the elevation plane and
change the viewing direction accordingly.
•
Other parameters: Frame rate, Frequency and Octave.
•
Angle
(on the Touch panel): display different rendering
views.
•
Layout
(on the Touch panel): display different layout
alternatives.
5. You may adjust
Num Beats
to change the number of heart
beats the acquisition should be based on.
6. Press
Multi beat
on the Touch panel.
The gated acquisition is started. Ask the patient to hold her/
his breath at end expiration. Keep the probe steady and look
for stitching artifacts in both the volume rendering and the
elevation plane in the lower left window of the screen, or use
Multi-Slice (page 6-10) to better assess the stitching quality.
Attention should be brought on stitching quality during
acquisition rather than volume rendering quality.
7. You may adjust
Angle
to get different rendering views. The
default top/down view is best for stitching quality
assessment.
8. Optimize the image quality as necessary (2D Gain, 4D Gain
and UD Clarity).
•
If desired, select a
Color map
to colorize the volume
rendering. The Depth color maps use a combination of
colors to improve depth perception.
Press
Red
or
Color Laser lines
on the Touch panel
(page 2) to visualize the 2D image locations in the volume
rendering. The 2D image locations are shown as overlaying
red or color coded lines following the surface in the volume
rendering. The color of the laser lines corresponds to the
color coding used for the 2D images.
Summary of Contents for Vivid S70
Page 8: ...i 6 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 62: ...Safety 2 30 Vivid S70 S60 User Manual BC092760 1EN 01 For China only...
Page 63: ...Device labels Vivid S70 S60 User Manual 2 31 BC092760 1EN 01...
Page 64: ...Safety 2 32 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 349: ...4D Auto MVQ Vivid S70 S60 User Manual 8 71 BC092760 1EN 01 Figure 8 37 Set Landmark stage...
Page 382: ...Measurements and Analysis 8 104 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 670: ...Probes 13 54 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 704: ...Maintenance 15 14 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 708: ...Index 4 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 709: ......