4D and Multi-plane Modes
6-10
Vivid S70 / S60
–
User Manual
BC092760-1EN
01
9. Press
Store
to store the acquisition.
NOTE:
It is recommended to acquire several heart cycles and use
Cycle select
to select the best one.
Figure 6-5. Stitching artifacts
Multi-Slice acquisition
Multi-Slice acquisition provides full volume data acquisition in
the same way as regular 4D single-beat or multi-beat
acquisition.
When using Multi-Slice acquisition the volume rendering display
is replaced by equidistant short axis views. The two Apical views
displayed on the left are for probe orientation purpose, while the
short axis views are used to ensure that the entire chamber is
included in the volume and to evaluate presence of stitching
artifacts (Figure 6-6).
1. While in 4D Live, press
Multi-Slice
on the Touch panel.
The
Multi Slice
screen is displayed showing equidistant
short axis views (Figure 6-6). The short axis views are
evenly distributed and maximized in size for best
assessment (e.g image quality, presence and visibility of all
walls, stitching artifacts when using real time multi beat
acquisition). Apical views are displayed on the left side for
orientation purpose.
Summary of Contents for Vivid S70
Page 8: ...i 6 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 62: ...Safety 2 30 Vivid S70 S60 User Manual BC092760 1EN 01 For China only...
Page 63: ...Device labels Vivid S70 S60 User Manual 2 31 BC092760 1EN 01...
Page 64: ...Safety 2 32 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 349: ...4D Auto MVQ Vivid S70 S60 User Manual 8 71 BC092760 1EN 01 Figure 8 37 Set Landmark stage...
Page 382: ...Measurements and Analysis 8 104 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 670: ...Probes 13 54 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 704: ...Maintenance 15 14 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 708: ...Index 4 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 709: ......