Image Optimization
5-34
Vivid S70 / S60
–
User Manual
BC092760-1EN
01
Using Strain
1. From TVI Mode, press
Strain
.
2. Adjust
Strain Start
close to the R-peak.
3. Adjust
Strain end
to end systole, typically near the T-wave.
4. Use the trackball to position the ROI frame over the area to
be examined.
5. Press
Select
. The instruction
Size
should be highlighted in
the trackball status bar.
NOTE:
If the trackball control Pointer is selected, press
Trackball
to
be able to select between Position and Size controls.
6. Use the trackball to adjust the dimension of the ROI.
Optimizing Strain
•
From an optimized Strain rate display adjust strain tracking
to pick out the systolic phase.
•
The main use of Strain is to map negative systolic
deformation. This means that
Strain start
and
Strain end
should be adjusted to pick out the systolic phase of the
cardiac cycle: Adjust
Strain start
close to the R-Peak.
Adjust
Strain end
to end systole, typically near the T-wave.
•
The maximum deformation that is color-coded can be
adjusted using the
Strain scale
control. If set too low, most
of the wall will show the color indicating maximum
deformation. If set too high, the maximum deformation color
is never reached.
•
Strain provides information only in the beam direction. The
apical view typically provides the best window since the
beams are then approximately aligned to the longitudinal
direction of the myocardium (except near the apex).
•
Low strain values may be masked out with a different color
using the
SI Reject
control.
Summary of Contents for Vivid S70
Page 8: ...i 6 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 62: ...Safety 2 30 Vivid S70 S60 User Manual BC092760 1EN 01 For China only...
Page 63: ...Device labels Vivid S70 S60 User Manual 2 31 BC092760 1EN 01...
Page 64: ...Safety 2 32 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 349: ...4D Auto MVQ Vivid S70 S60 User Manual 8 71 BC092760 1EN 01 Figure 8 37 Set Landmark stage...
Page 382: ...Measurements and Analysis 8 104 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 670: ...Probes 13 54 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 704: ...Maintenance 15 14 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 708: ...Index 4 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 709: ......