Image controls
Vivid S70 / S60
–
User Manual
5-49
BC092760-1EN
01
Image controls
Control panel
Touch panel and rotaries
2D Gain
2D
When rotated clockwise, increases the overall gain applied to the received echo
signals equally for all depth.
Time Gain Compensation TGC
2D
Compensates for depth-related attenuation in an image. The sliders nearest the
operator affect the far field. TGC amplifies returning signals to correct for the
attenuation caused by tissue at increasing depths.
Auto
2D
Turns
Auto Tissue
setting on/off (see ‘Auto Tissue’ on
).
Depth
2D
Sets the maximum (far field) distance that will be imaged. Decreasing the depth
may allow higher frame rates.
Line Density
2D (non-cardiac)
Controls the spatial resolution of the image versus the image update rate.
HD (High Definition)
2D
Smoothens the speckle and reduces the noise in the image.
Width/2D width
2D, CF, TVI, TT, SRI,
SI, TSI
Controls the size or angular width of the 2D image. A smaller angle generally
produces an image with a higher frame rate.
Summary of Contents for Vivid S70
Page 8: ...i 6 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 62: ...Safety 2 30 Vivid S70 S60 User Manual BC092760 1EN 01 For China only...
Page 63: ...Device labels Vivid S70 S60 User Manual 2 31 BC092760 1EN 01...
Page 64: ...Safety 2 32 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 349: ...4D Auto MVQ Vivid S70 S60 User Manual 8 71 BC092760 1EN 01 Figure 8 37 Set Landmark stage...
Page 382: ...Measurements and Analysis 8 104 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 670: ...Probes 13 54 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 704: ...Maintenance 15 14 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 708: ...Index 4 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 709: ......