Customize Configuration
12-52
Vivid S70 / S60
–
User Manual
BC092760-1EN
01
NOTE:
Only jobs that failed or are in hold can be resent.
2. Press
Resend
.
To send a job to a new destination:
This is typically used if the configuration of the server is
changed.
1. Select the job(s) to send in the
DICOM job spooler
window.
NOTE:
Only jobs that failed or are in hold can be sent to a new
destination.
2. Press
Send to...
.
A dialogue window is displayed.
3. Select the new destination from the
Destination
pull-down
menu.
4. Press
Send
.
To hold a job:
1. Select the job(s) to hold in the
DICOM job spooler
window.
NOTE:
Only inactive jobs can be set on hold.
2. Press
Hold
.
3. To undo hold, press
Resend
.
DICOM images
To configure DICOM images:
1. Press
Utility
/
Config
on the Touch panel and log on as
administrator.
2. Select the
Connectivity
category and
DICOM
subgroup.
The
DICOM
sheet is displayed.
Figure 12-36. The DICOM sheet
Summary of Contents for Vivid S70
Page 8: ...i 6 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 62: ...Safety 2 30 Vivid S70 S60 User Manual BC092760 1EN 01 For China only...
Page 63: ...Device labels Vivid S70 S60 User Manual 2 31 BC092760 1EN 01...
Page 64: ...Safety 2 32 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 349: ...4D Auto MVQ Vivid S70 S60 User Manual 8 71 BC092760 1EN 01 Figure 8 37 Set Landmark stage...
Page 382: ...Measurements and Analysis 8 104 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 670: ...Probes 13 54 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 704: ...Maintenance 15 14 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 708: ...Index 4 Vivid S70 S60 User Manual BC092760 1EN 01...
Page 709: ......