B30 Patient Monitor
3-24
Document no
.
2044677-001
Check that the HR value is calculated from P1, when ECG is not measured (ECG cable
disconnected).
"
SpO
2
measurement
18. SpO
2
probe detection
Check that the message ‘No probe’ is shown, when no SpO
2
sensor is connected to the
module. Connect an SpO
2
finger probe to the module (with the interconnection cable, if
needed). Check that the message ‘Probe off’ is shown when the probe is not connected to
a finger.
"
19. Test measurement
Connect the SpO
2
probe onto your finger. Check that the reading of 95-99 and SpO
2
waveform appears. Check that the HR value is calculated from SpO
2
when ECG and InvBP
(P1) are not measured.
NOTE: a functional tester cannot be used to assess the accuracy of a pulse oximeter probe or a
pulse oximeter monitor
"
Non Invasive Blood Pressure measurement
20. Communication and memories
Enter the NIBP module service menu:
Parameters - NIBP
Check that the Time-outs, Bad checksums and Bad c-s by mod values are not increasing
faster than by 5 per second. Check also that the NIBP board memories have passed the
internal memory test, i.e. the RAM, ROM and EEPROM show all OK.
"
21. Pump and valves
Check the pump and valves.
Select
Pneumatics
from the NIBP menu. Connect a pressure manometer to the NIBP
module cuff connector.
Select
Start Pump
and press the ComWheel. Check that the pump turns on and the
pressure inside the tubing system starts to increase. Stop the pump by pressing the
ComWheel again when the pressure reaches 280 mmHg.
Select
Open Exh2
. Press the ComWheel and check that the pressure inside the tubing
system starts to drop, then press the ComWheel again. If necessary, turn the pump on
again for a moment to increase the pressure inside the tubing system.
Select
Set Valve
. Press the ComWheel and set the value under the text Pulse Valve to
number 150 by turning the ComWheel. Press the ComWheel again and check that the
pressure inside the tubing system starts to drop. Finish the test by selecting
Previous
Menu
.
"
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......