![background image](http://html1.mh-extra.com/html/ge/b30/b30_technical-reference-manual_4013170152.webp)
B30 Patient Monitor
5-26
Document no
.
2044677-001
4.4.5 NIBP Pneumatics
How to check Interval 20 mmHg -> 185 mmHg
Select
Start pump
at different combinations of the valves open/closed and push the
ComWheel. The module counts the time it takes for the pressure to go up from 20 mmHg to
185 mmHg and displays it. When all the valves are closed, the pump should be able to pump
the pressure in about 1 to 4 seconds into an adult cuff wrapped around a pipe. The pump does
not stop without selecting
Stop Pump
by pushing the ComWheel.
NOTE: To redo the test, must go back to the previous menu.
Start Pump/Stop Pump
:
A manual control for the
pump. The selection changes to
Stop Pump
when the
pump turns on.
Open Exh1/Close Exh1
:
This function is not in use
with E-PSM(P)W modules.
Open Exh2/Close Exh2
:
A manual control for the
exhaust valve 2. The selection changes to
Close Exh2
when the valve is opened.
Open Zero valve
:
This function is not in use with
E-PSM(P)W modules.
Set Valve
:
The opening of the pulse valve is adjusted
between 0 and 255 (0 for fully closed and 255 for fully
open). First push the ComWheel, then turn it to adjust
the value on the screen and finally push to set the
value.
Reset Clock
is not in use with E-PSM(P)W modules.
Pneumatics Data field
For information on general items
Pressure
,
Zero
,
Protect handle
,
Calibr. prot.
,
+15 V power
,
AD0
to
AD7
as well as
Timeouts
etc.,
Pump
,
Exh1 Valve
, and
Exh2 Valve
show their states.
NOTE:
Exh1 Valve
has no effect on the module.
Pulse Valve
shows how much the valve is opened (0 to 255) during Valve Setting.
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......