Adjustments and calibrations
NOTE: Use only properly maintained, calibrated and traceable measurement equipment for
the specified calibrations and adjustments to ensure accuracy.
5.1 NIBP calibrations
The electronics of the NIBP pressure measurement is calibrated at the factory. The processor
automatically maintains the zeroing pressure. If the zero point of the pressure transducer drifts
more than specified, an error message is given and the NIBP board should be recalibrated or
replaced.
Recalibrate the NIBP measurement once a year. The checking and recalibration can be done in
the NIBP service menu.
The calibration of the primary pressure channel can also be checked from the NIBP setup
menu (
NIBP - NIBP Setup - Calibration Check
). In this case, the auto zeroing is performed at
start - remove the hose before entering to ensure atmospheric pressure to the pressure
transducers - the primary pressure is displayed. The zero-offset value should then be zero.
Check the intake air filter as part of the calibration check. Change the filter if it is visibly dirty.
Calibration check
1.
Enter
Calibration
menu:
Monitor Setup -
Install/Service
(password 16-4-34) -
Service
(26-23-8) -
Parameters -
NIBP - Calibrations
2.
Select
Calibration Check
and push the ComWheel.
3.
Connect an external precision manometer to the module.
4.
Pump the following pressures to manometer and check the difference between the
manometer and monitor pressure display (The zeroing offset is automatically subtracted
from the pressure readings).
PR
ES
TN_cali
b
_menu.vsd
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......