Service procedures
8-13
Document no. 2044677-001
Remove any paper chaff from the paper compartment.
Clean the thermal printhead and the small glass window in front of the static brush with a
cotton swab dipped in isopropyl alcohol. Avoid contact with the rubber paper roller.
NOTE: Be careful to limit the application of alcohol to the thermal printhead and the window.
Leave the paper compartment empty and close the hatch.
"
2.
Check external parts
•
all connectors are intact and attached properly
•
the module box and the locking system are intact
"
Reattach the module and check that the locking system moves properly.
CO
2
measurement
3.
Check Mini D-Fend
Detach the Mini D-fend. Check the condition of the rubber O-rings on the metal Mini D-
fend connectors, located in the module front cover.
If necessary, detach the connectors by first disconnecting the tubes, then removing the
locking rings from the back of the front cover.
"
Replace the Mini D-fend and sampling line with new ones.
NOTE: Use only Datex-Ohmeda sampling lines in order to ensure proper functioning.
Connect the Extension Module, N-FC or N-FCREC to the monitor.
Turn on the monitor.
Configure the monitor screen so that the CO
2
curve is shown, for example as follows:
Monitor Setup
- Screen 1 Setup - Waveform fields -
Field 6 - CO2
Digit Fields
Lower Field 1 - Gases
4.
Module software
Wait until the message ‘Calibrating gas sensor’ disappears from the screen, then enter
the service menu.
Monitor Setup
-
Install/Service
(password 16-4-34)
- Service
(password 26-23-8).
Write down the information regarding the ExtensionModule software.
5.
Module bus communication
Enter the service menu
General: Parameters - Gas Unit - General
Check that the Timeouts, Bad checksums and Bad c-s by mod values are not increasing
faster than by 5 per second.
If one of the values is increasing faster, it indicates a failure in module bus
communication.
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......