B30 Patient Monitor
1-10
Document no
.
2044677-001
•
To avoid explosion hazard, do not use the monitor in presence of flammable
anesthetics.The monitor measures only non-flammable anesthetics.
•
Do not touch the patient, table, instruments, modules or the monitor during defibrillation.
External connection
•
Do not connect any external devices to the monitor other than those specified.
Fuse replacement
•
Replace a fuse only with one of the same type and rating.
Explosion hazard
•
To avoid explosion hazard do not use the monitor in the presence of flammable
anesthetics.
•
Do not incinerate a battery or store at high temperatures as it will explode.
Patient safety
•
Do not perform any testing or maintenance on the monitor while it is being used on a
patient.
•
PACEMAKER PATIENTS: The impedance respiration measurement may cause rate
changes in Minute Ventilation Rate Responsive Pacemakers. In this case, set the
pacemaker rate responsive mode off or turn the monitor impedance respiration
measurement off.
•
Never install the monitor so that it is above the patient.
•
The monitor must not be used without manufacturer approved mounting attached.
•
Operation of the monitor outside the specified values may cause inaccurate results.
Cleaning and service
•
Only trained personnel with proper tools and test equipment should perform the tests
and repairs described in this manual. Unauthorized service may void the monitor
warranty.
•
Always unplug the monitor before cleaning or service. After cleaning or service ensure
that every part of the monitor is dry before reconnecting it to the power supply.
•
Do not touch any exposed wire or conductive surface while any cover is removed and the
monitor is energized. The voltages present can cause injury or death.
•
Pins of connectors identified with the ESD warning symbol should not be touched.
Connections should not be made to these connectors unless ESD precautionary
procedures are used. For details, see section “
1.2.2. ESD precautionary procedures
”.
•
NOTE! The monitor is always internally powered when the batteries are connected.
•
Always perform an electrical safety check and a leakage current test on the monitor after
service.
•
Handle the water trap and its contents as you would any body fluid. Infectous hazard
may be present.
Accessories
•
Use only accessories, including mounts and batteries, and defibrillator-proof cables and
invasive pressure transducers approved by GE Healthcare. For a list of approved supplies
and accessories, see the “Supplies and Accessories” catalog delivered with the monitor.
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......