Electrical Safety Tests
3-5
Document no. 2044677-001
•
Failure of the power cord strain relief is very common. Often times users of the equipment
pull on the power cord itself, rather than the power cord plug, to unplug the patient
monitor from a wall receptacle. If in doubt, test for continuity through each conductor of
the power cord connector and plug.
•
Verify line, neutral, and earth conductors are properly connected to the power cord plug
and are not short-circuited. Replace the power cord, as necessary with a
regulatory-approved cord for the country of use.
WARNING
Use only AC power cords recommended or manufactured by GE.
2.6 Ground (Earth) Integrity
Listed below are two methods for checking the ground (earth) integrity, “Ground Continuity
Test” and “Impedance of Protective Earth Connection.” These tests determine whether the
device's exposed metal and power inlet's earth (ground) connection has a power ground fault
condition
Perform the in accordance with your local regulations.
2.6.1 Ground Continuity Test
Refer to the instructions contained with the safety analyzer to perform each test.
The measuring device (MD) in the diagram below may be a DMM or part of a safety analyzer.
NOTE: The measuring device (MD) represents the network and voltage measuring instrument
and its frequency characteristics per IEC 60601-1.
2.6.2 Impedance of Protective Earth Connection
This test, unlike a ground continuity test, will also stress the ground system by using special
ground bond testers.
This test normally is only required as a manufacturing production test to receive safety agency
compliance. Some country agencies do require this test after field equipment repairs (e.g.,
Germany’s DIN VDE 0751 standards). Consult your country/local safety agency if in question.
Compliance is checked by the following steps:
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......