B30 Patient Monitor
4-8
Document no
.
2044677-001
3
E-PSM(P)W Troubleshooting
3.1 Troubleshooting flowcharts
3.1.1 Troubleshooting for NIBP parameter
Insert new module and turn
power on.
Does fault still appear?
Module ID on screen?
Does another
module work in
same slot?
Error message on
screen?
See error code explanation in
service manual and fix it.
Do they work?
Check tubes and
connectors.
Find leak and fix it.
Leak test in Service
Menu OK?
Pump check in
Service Menu OK?
Key check in
Service Menu OK?
Check keyboard
connector. If
OK, change
front panel.
OK?
NO
YES
YES
NO
NO
NO
YES
NO
NO
YES
YES
YES
NO
Check pump
connector.
If OK, change
pump.
YES
START
NO
OK?
Replace
the module.
Check and
replace
NIBP board.
NO
PS
M
_
mo
dule_
tr
ou
bles_f
lo
w
ch.vsd
NO
Fault not in original
module
Start NIBP
without hose
Cuff loose
message appears
on screen?
Check and replace
- STP Input Board
- NIBP flex board
- connectors
YES
NO
Cuff ID OK but measurement
does not work
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......