E-PSM(P)W module introduction
1-43
Document no. 2044677-001
Table 6
Temp connector (T1, T2)
4.6 Measurement principle
4.6.1 NIBP
NIBP (Non-Invasive Blood Pressure) is an indirect method for measuring blood pressure.
The NIBP parameter conforms to EN 1060-1:1995 Specification for Non-invasive
sphygmomanometers.
The NIBP measurement is performed according to the oscillometric measuring principle. The
cuff is inflated with a pressure slightly higher than the presumed systolic pressure, and
deflated at a speed based on the patient’s pulse, collecting data from the oscillations caused
by the pulsating artery. Based on these oscillations, values for systolic, mean, and diastolic
pressures are calculated.
The following parts are necessary for the NIBP measurement:
•
E-PSM(P)W module
•
twin hose (adult or infant model)
•
blood pressure cuffs (various sizes)
4.6.2 ECG
Electrocardiography analyzes the electrical activity of the heart by measuring the electrical
potential produced with electrodes placed on the surface of the body.
ECG reflects:
•
electrical activity of the heart
•
normal/abnormal function of the heart
Temp connector
Pin No
Signal
1
Sensor drive current
2
Input from temperature sensor, channel 1
3
Not connected
4
Not connected
5
Thermistor ID (LOW= Temperature error,
HIGH=YSI 400 series)
6
Cable shield
7
Analog ground
8
Input from temperature sensor, channel 2
9
Not connected
10
Not connected
11
Digital ground
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......