B30 Patient Monitor
4-12
Document no
.
2044677-001
‘Cuff occlusion’ message
1. Cuff and/or hose occluded.
Reason:
- cuff tube kinked
- Straighten tube.
- tubes inside the shield kinked
- Straighten tubes.
- tubes inside module kinked
- Straighten tubes.
- occlusion inside/outside module - Remove occlusion.
2. Cuff, hose, and tubes OK.
Reason:
- fault in pressure transducer
- Replace the NIBP board.
- fault in A/D converter
- Replace the NIBP board.
- faulty calibration
- Check calibration.
‘Calibration switch on’
message
EEPROM protection has been
handled by pressing module buttons
Auto ON/OFF
and
Start/Cancel
simultaneously for 3 seconds.
Enables setting the protection OFF in the
Calibration
menu. Press the buttons
again if you are not going to calibrate.
‘Calibration not protected’
message.
Calibration protection is set to OFF.
Set the protection ON in the NIBP
Calibration
menu.
Problem
Cause
What to do
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......