B30 Patient Monitor
8-30
Document no
.
2044677-001
7.2.1 CO
2
measurement
‘REPLACE D-FEND’ message
Amb – Work pressure difference too big.
Indicates residue build-up on the water trap membrane. This decreases
air flow. Replace the D-fend.
‘SAMPLE LINE BLOCKED’ message Amb – Work pressure difference too big.
Sampling line or water trap is occluded. Water trap container is full. If
occlusion persists, check internal tubing for blockages.
No response to any gas
Check Sample Gas Out.
Amb - Work pressure low, flow too small and pump voltage too high.
Sampling line, water trap, or internal tubing is blocked or loose, or
improperly attached. Gas out connector or tubing is blocked.
Zero valve malfunction. Pump failure or pump is worn. Supply voltage
missing. Serial communication error.
Sudden increase in gas display
Water trap malfunction. Check all internal tubing and the interior of the
water trap for occlusions or leaks. Replace water trap. Check flow rates.
Abnormally high (or abnormally
low response to CO
2
or sudden occlusion warning
Pressure transducer failure. Check the Ambient and Amb – Work
pressures in
Gases
service menu.
Strong drift in all gases
Leak in sampling line or internal tubing (especially in conjunction with too
low readings).
Problem
Cause
What to do
Action
too low ETCO
2
value
•
sudden decrease in circulation
•
pulmonary embolism
•
hyperventilation
•
very large dead-space
•
large shunting
•
leak in sampling system
•
calibration error
•
high by-pass flow from ventilator
•
check all connections
•
check calibration
too high ETCO
2
•
hypoventilation
•
increased metabolism
•
D-fend contaminated
•
calibration error
•
change D-fend
•
check calibration
waveform clipped
•
incorrect scaling
•
change scale
Problem
Cause/What to do
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......