![GE b30 Technical Reference Manual Download Page 220](http://html1.mh-extra.com/html/ge/b30/b30_technical-reference-manual_4013170220.webp)
B30 Patient Monitor
8-20
Document no
.
2044677-001
NOTE: The finger of the insulator plate secures the
interconnection flex connector. While reassembling
the insulator plate ensure that the finger supports the
connector correctly.
9. Disconnect the Nafion and silicon tubes from the
mini D-fend connectors.
NOTE: While reassembling the nafion tube ensure the
correct routing for both modules. If the nafion tube is
not correctly routed it may get bent or jammed when
reassembling the cover. Bent or pinched tube can
get permanently blocked. The block may appear
later when the tube warms up.
Correct nafion tube routing in
N-FCREC
module, arrow
pointing at the mini D-fend connectors.
Correct nafion tube routing in
N-FC
module, arrow
pointing at the mini D-fend connectors.
10. Disconnect the gas out tube from the pump unit
and remove the front panel unit.
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......