N-FCREC Module disassembly and reassembly
8-23
Document no. 2044677-001
5.1 Adjustments and calibrations
5.1.1 Calibrating
The airway module should be calibrated once every six months or whenever there are
indications of errors in the gas readings.
Calibrate the gas measurement with the Datex-Ohmeda calibration gas. Do not use any other
calibration gases.
−
Use the regulator 755534.
−
Use gas 755580 only and set the FiO2 level in the CO2 setup menu to 21-40%
Airway Gas
- CO2 setup.
.O
2
concentration to 20%.
Use only recommended calibration gases to guarantee a successful calibration.
During gas calibration, % units are always used for CO
2
regardless of selected measuring units.
NOTE: Ensure that the calibration gas and regulator are functioning properly before calibration.
Perform the annual maintenance on the regulator as required.
Figure 8
Attaching regulator to the calibration can
1.
Attach the regulator to the gas container.
2.
Attach a new sampling line to the water trap. Connect the loose end of the sampling line
to the regulator on the gas container.
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......