B30 Patient Monitor
5-22
Document no
.
2044677-001
NOTE: Messages ‘Zeroing’ and ‘Zeroed’ is shown in the NIBP message field and next to the
Calibrate
selection momentarily. After this, a pressure bar will appear beside the menu and the
text ‘Calibrating’ is shown in the NIBP digit field.
4.
Calibration:
−
Connect the NIBP hose to the module connector.
−
Connect an external manometer with a pump to both tubes of the hose.
−
Pump about 200 mmHg pressure.
−
Verify that both pressure values, B1 and B2, shown in the prompt field of the
calibration menu match the manometer reading. If not, adjust the by turning the
ComWheel.
−
Press
ComWheel
to complete the calibration.
NOTE: Messages ‘Calibrating” and ‘Calibrated’ are shown in the NIBP message field and next to
the
Calibrate
selection.
NOTE: When calibrating NIBP, always change the displayed pressure value slightly with the
ComWheel, even in cases where the value would be correct. For example, change the value
one step higher and then back one step lower. The ‘Calibrated’ text should appear in the
display. This ensures that the calibration procedure is correctly registered and stored by the
module.
5.
Change the
Protection
setting from OFF to ON to disable
Calibrate
selection - the color of
the
Calibrate
selection changes from white to grey.
6.
Press the NIBP module buttons
Auto ON/OFF
and
Start Cancel
simultaneously for 3
seconds. This disables the
Protection
selection - the color of the
Protection
selection
changes from white to grey.
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......