B30 Patient Monitor
1-40
Document no
.
2044677-001
Figure 24
Serial communication of E-PSM(P)W module
Signals and isolation barrier
The communication signals transfer over the isolation barrier by using high isolation voltage
(6kV) opto isolators.
Power supply section
The power for the electronics on the floating part of the STP and the ECG boards is made on
each board with the switching power supplies connected to a high voltage isolated
transformer. The switching power supplies on the STP and ECG boards are synchronized to the
frequency, about 340kHz of the switching power supply on the NIBP board. The NIBP board
supplies non-isolated 5 V to the ECG and STP boards. The module uses only Vmod 13.8 - 16 V
voltage of the frame. The other voltages of the measuring boards are made by the switching
power supplies and regulators or the linear regulators. Each measuring board is protected
against overloading with PTC type automatic fuses.
RS 485
DRIVER
DATA
NDATA
RECEIVE DATA
SEND DATA
SEND/RECEIVE
N
IB
P
B
O
A
R
D
C
O
N
T
R
O
LL
E
R
S
NIBP BOARD
RS 485
DRIVER
S
T
P
B
O
A
R
D
C
O
N
N
E
C
T
O
R
RECEIVE DATA
SEND DATA
SEND/RECEIVE
S
T
P
B
O
A
R
D
C
O
N
T
R
O
L
LE
R
STP BOARD
PAT
IENT IS
OLA
T
ION
RS 485
DRIVER
E
C
G
B
O
A
R
D
C
O
N
N
E
C
T
O
R
DATA
NDATA
E
C
G
B
O
A
R
D
C
O
N
T
R
O
LL
E
R
ECG BOARD
P
A
TIENT
ISOL
ATION
DIRECT ECG (PWM)
PS
M_se
rial_co
mmuni
cation
.vsd
Mo
dule b
u
s con
nec
to
r
DATA
NDATA
sy
nchron
izatio
n sign
al
+5
V
+5 V
+5 V
RECEIVE DATA
SEND DATA
SEND/RECEIVE
NIB
P
BOAR
D CON
N
ECT
O
R
Summary of Contents for b30
Page 2: ......
Page 12: ...B30 Patient Monitor 8 Document no 2044677 001...
Page 13: ...1 Introduction...
Page 14: ......
Page 61: ...2 Installation...
Page 69: ...3 Maintenance...
Page 70: ......
Page 105: ...4 Troubleshooting...
Page 106: ......
Page 125: ...5 Service Menu...
Page 126: ......
Page 158: ...B30 Patient Monitor 5 32 Document no 2044677 001...
Page 159: ...6 Field replaceable unit...
Page 160: ......
Page 166: ...B30 Patient Monitor 6 6 Document no 2044677 001 1 6 Frame for B30 Frame for B30...
Page 183: ...Frame disassembly 6 23 Document no 2044677 001...
Page 190: ...B30 Patient Monitor 6 30 Document no 2044677 001...
Page 191: ...7 Technical specification...
Page 192: ......
Page 198: ...B30 Patient Monitor 7 6 Document no 2044677 001...
Page 199: ...8 N F C REC Module...
Page 200: ......
Page 238: ...B30 Patient Monitor A 6 Document no 2044677 001...
Page 251: ......